ログイン
言語:

WEKO3

  • トップ
  • ランキング
To
lat lon distance
To

Field does not validate



インデックスリンク

インデックスツリー

メールアドレスを入力してください。

WEKO

One fine body…

WEKO

One fine body…

アイテム

  1. 紀要類
  2. 横浜市立大学紀要(1951-)
  3. 自然科学系
  4. 体力医学編(1972-)
  5. Vol.17(1989)

スポーツテスターPE-3000の信頼性について

https://ycu.repo.nii.ac.jp/records/804
https://ycu.repo.nii.ac.jp/records/804
869ad0a3-2b3d-4272-84a8-e33bcb06ecfe
名前 / ファイル ライセンス アクション
KJ00000160640.pdf KJ00000160640.pdf (641.4 kB)
Item type [ELS]紀要論文 / Departmental Bulletin Paper(1)
公開日 2017-04-21
タイトル
タイトル スポーツテスターPE-3000の信頼性について
タイトル
タイトル Validity of heart rates obtained by the Sports tester PE-3000 during several exercise conditions
言語 en
言語
言語 jpn
資源タイプ
資源タイプ識別子 http://purl.org/coar/resource_type/c_6501
資源タイプ departmental bulletin paper
雑誌書誌ID
収録物識別子タイプ NCID
収録物識別子 AN00246981
論文名よみ
タイトル 〓
著者 玉木, 伸和

× 玉木, 伸和

WEKO 2053

玉木, 伸和

ja-Kana タマキ, ノブカズ

Search repository
TAMAKI, Nobukazu

× TAMAKI, Nobukazu

WEKO 2054

en TAMAKI, Nobukazu

Search repository
著者所属(日)
値 横浜市立大学体育医学教室
著者所属(英)
言語 en
値 Institute of Sport Medicine, Yokohama City University
抄録(日)
内容記述タイプ Other
内容記述 無線搬送式で小型・軽量タイプの心拍数モニターであるスポーツテスターPE-3000の信頼性を, 種々の運動負荷のもとで検討し, 以下の結果を得た。1)スポーツテスターの60秒モードで記憶した心拍数は, 心拍数の変化幅の大きい運動初期や運動直後に, 実際の心拍数と比較して過大評価あるいは過小評価する傾向が認められた。この傾向は, 運動強度が強くなるほど増大し, 特に1分毎の激しい間欠運動では顕著となった。2)15秒モードで得た心拍数は, 実際の心拍数変化に比較的近い値を示したが, 過大評価や過小評価の傾向はわずかに認められた。3)5秒モードによる結果は, 運動負荷の形態が異なっても, また強度の増減が激しい場合においても, 実際の心拍数と同じ値を示した。4)このような結果の差異は, 記憶時間内の平均心拍数がメモリーされるのではなく, 常にその時点での最新の心拍数が記憶されることが原因であることが明らかとなった。以上のことから, 忠実に運動時の心拍数変化を記録することが必要な場合には, より短い記憶モードを使用すべきである。しかし, 長時間の心拍応答の変化パターンが明らかになればよいという場合には, 60秒モードの使用も有効であろう。いずれにしても, 特性と限界を認識し, 目的に応じて使用するならば, スポーツテスターPE-3000は, 心拍数モニターとして非常に有益な装置である。
抄録(英)
内容記述タイプ Other
内容記述 The validity of heart rates obtained by the Sports tester PE-3000 was assessed by comparing the measuring values with simultaneous ECG recordings. One subject performed several exercise tests on a bicycle ergometer. Heart rates recorded before, during and after exercise. Under the conditions of increasing heart rate, heart rates recorded by the Sport tester PE-3000's 60-seconds-mode were frequently higher than those measured at the same time by ECG. While, under the conditions of decreasing heart rate, heart rates obtained by the Sport tester were lower than those of ECG. This tendency was increase with increasing work load. However, heart rates recorded by the Sport tester's 15-seconds-mode were closely related to those measured by ECG. In this condition, heart rates differed at most by only 3 beats min^<-1>. Furthermore, heart rates recorded by the Sport tester's 5-seconds-mode were equal to those measured by ECG. These differences were caused mainly by the memory methods of the Sport tester PE-3000. Therefore, the shorter memory mode is suitable for measuring accurate heart rate and analyzing intensity of exercise.
書誌情報 横浜市立大学紀要. 体力医学編
en : The journal of Yokohama City University. Series of sport sciences and medicine

巻 17, p. 1-7, 発行日 1989-01-26
表示順
内容記述タイプ Other
内容記述 3
アクセション番号
内容記述タイプ Other
内容記述 KJ00000160640
ISSN
収録物識別子タイプ ISSN
収録物識別子 09115277
戻る
0
views
See details
Views

Versions

Ver.1 2023-06-19 10:01:11.294227
Show All versions

Share

Mendeley Twitter Facebook Print Addthis

Cite as

エクスポート

OAI-PMH
  • OAI-PMH JPCOAR 2.0
  • OAI-PMH JPCOAR 1.0
  • OAI-PMH DublinCore
  • OAI-PMH DDI
Other Formats
  • JSON
  • BIBTEX

Confirm


Powered by WEKO3


Powered by WEKO3